26
RS-232C
Transmission system : EIA RS-232C
Transmission form : Asynchronous, bi-directional, half duplex
Data format : Baud rate : 600, 1200, 2400, 4800, 9600, 19200 bps
Data bits : 7 bits or 8 bits
Parity bit : EVEN, ODD (Data bits 7 bits)
NONE (Data bits 8 bits)
Stop bits : 1 bit or 2 bits
Code : ASCII
Terminator : <CR> or <CR><LF>
Circuit
Connection to the weighing instrument Connection to an external device
D-Sub 9-pin (BALANCE/SCALE) DIN 8-pin (COMP.OUT)
Pin
No.
Signal
name
Direction Description
Pin
No.
Signal
name
Description
1 ⎯ ⎯ N.C. 1 HI HI contact-output
2 RXD Input Receive data 2 COM COM contact-output
3 TXD Output Transmit data 3 TXD Transmit data (RS-232C output)
4 ⎯ ⎯ N.C. 4 LO LO contact-output
5 SG ⎯ Signal ground 5 SG Signal ground
6 DSR Input Data set ready 6 OK OK contact-output
7 RTS Output Request to send 7 DSR Data set ready (RS-232C output)
8 ⎯ ⎯ N.C. 8 RXD Receive data (RS-232C output)
9 ⎯ ⎯ N.C.
(AD-8922A is a DTE. Connects to a DCE
such as a weighing instrument using a
straight through cable.)
9876
SG
SG
TXD
RXD
HI
OK
LO
COM
HI
OK
LO
DSR
AD-8922 A interior
AD-8922A interior
TXD
RXD
6
1
4
2
5
3
7
8
54321
To a weighing instrument
To an external device such as
a personal computer or a printer
Inch screw thread
(#4-40UNC)