English
GB 69
Display Error code Description of the error
EC, E06 Format error
A command includes incorrect data.
Confirm command.
ex. Data is numerically incorrect.
EC, E07 Range error for a parameter
The received data exceeds the range that the balance can
accept.
Confirm parameter range of command.
Internal condition information
There is no problem when this is displayed for a few
seconds and returns to weighing mode.
Turn the balance off then on again, if this display is
continuously displayed.
EC, E11 Stability error
The balance can not stabilize due to an environmental
problem. The balance is able to inter the weighing mode
with the CAL command.
Prevent vibration, draughts, temperature changes, static
electricity and magnetic fields.
EC, E16 Internal weight error
This is a calibration error.
Confirm that there is nothing on the pan and retry the
calibration or calibration test.
EC, E17 Internal weight error
This is a calibration error.
Retry the calibration or calibration test.
EC, E20 Calibration error
The calibration weight is too heavy. The balance is able to
enter the weighing mode with the CAL command.
EC, E21 Calibration error
The calibration weight is too light. The balance is able to
enter the weighing mode with the CAL command.
Over load
This is a warning that a weight beyond the balance
capacity has been placed on the pan.
Remove the weight from the pan.