English
GB 25
4.1. Response Adjustment
This function stabilises the weight value, helping to reduce the effects caused by draughts and/or vibra-
tion at the balance location. This function makes adjustments by automatically analysing the environment
or by manual operation. The function has three stages:
Indicator Function table Summaries
FAST Cond 0 Fast response, Sensitive value
MID. Cond 1
SLOW Cond 2 Slow response, Stable value
Response Indicator
If the automatic response adjustment is causing problems – see Section “Manual Response Adjustment”.
The response adjustment can be changed at “Condition (Cond)” of “Environment & Display (bASFnc)” in
the function table. Refer to “8. Function table”.
4.2. Automatic Response Adjustment
RESPONSE
.
.
Press and hold the key
This automatically updates the response adjustment by analysing the influence of the environment on the
weight data.
Operation
1. Press and hold the [MODE] key until RESPONSE is displayed.
2. The balance analyses and and updates the response adjustment.
If you want to cancel this update, press the [CAL] key.
Note: Do not allow vibration or draughts to affect the balance.
3. The balance returns to the weighing mode automatically and displays
the updated response indicator.
4. Adapting to the Environment
Release the key