A&D GH Series Manual de Instruções Página 18

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 71
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 17
A&D GR Series Instruction Manual
GB 18
Operation
1. Connect the AC adapter and warm up the balance for
at least one hour with nothing on the pan.
2. Press and hold the [CAL] key until the balance displays
CC in and then release the key.
3. The balance measures the zero point.
Prevent vibration and draughts from affecting the balance.
4. The measured zero point data is displayed.
5. Ready for the internal weight measurement.
6. The balance measures the internal weight.
Prevent vibration and draughts from affecting the balance.
7. The internal weight data is displayed.
8. The balance informs you when the calibration test is
finished. If the “GLP output (inFo)” of the “Function Table”
is set to “/” or “2”, “Calibration Test Report” is output by
the RS-232C interface.
9. The balance will automatically return to the weighing mode
after the calibration test is finished.
Press and hold the key
Release the key
GLP output
Calibration Test Report
3.4. Calibration Test Using the Internal Weight
Vista de página 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 70 71

Comentários a estes Manuais

Sem comentários