A&D GR Series Instruction Manual
GB 68
• Do not disassemble the balance. Contact your local A&D dealer if your balance needs service or repair.
• Please use the original shipping box for transportation.
• Do not use organic solvents to clean the balance. Use a warm lint free cloth that is damp, along with a
detergent for cleaning.
• The “Floor Plate of the Weighing Chamber” can be removed and cleaned.
• See “Section 1.5. Caution” when operating the balance.
17.1. Error Codes
Display Error code Description of the error
EC, E00 Communications error
A protocol error occurred in communications.
Confirm the format, baud rate and parity.
EC, E01 Undefined command error
An undefined command was received.
Confirm command.
EC, E02 Not ready
A received command can not be processed.
Adjust the delay time to transmit the command.
ex. The balance received a Q command, but not in the
weighing mode.
ex. The balance received a Q command while processing
a RE-ZERO command.
EC, E03 Time over error
If t-Up 1 of the function list is set, the balance did not
receive the next character of a command within the time
limit of one second.
Confirm communication.
EC, E04 Excess characters error
The balance received excessive characters in a command.
Confirm command.