A&D OP-05 Especificações Página 47

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 90
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 46
45
CHECK WEIGHING MODE 2 (WEIGHING MODE CALF-14 =
6)
Output
Terminal
Output Condition
Zero band
Gross Zero band
Hi-Hi Net > Hi-Hi Limit
Hi
Target Weight + Hi-Hi Limit Net > Target Weight
Go
Target Weight + Hi Limit Net Target Weight – Lo Limit
Lo
Target Weight – Lo-Lo Limit Net < Target Weight – Lo Limit
Lo-Lo Net < Lo-Lo Limit
When an output condition is
established, a relevant output
terminal is turned on (power
continuity with output COM).
An output terminal number can
be selected with the general
functions, OUTF-01 through
OUTF-08.
The setpoint signal output is
turned off unconditionally when
the operation mode is other
than the normal mode.
Vista de página 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 89 90

Comentários a estes Manuais

Sem comentários