A&D AD-4430B Especificações Página 45

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 56
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 44
AD-4430B - 43 -
7.6. The Parameter List For The Function List
When performing maintenance, use the following list as a memorandum.
When making inquiries about the product, inform your local A&D dealer of the user settings.
7.6.1. Basic Functions
Function No.
Parameter
and range
Function
name
Description
Default
value
User
setting
MEMO
Fnc01
0000 to 1111
Disable key
switch
Each digit of the setting corresponds to a key
switch. Only available in the weighing mode.
0 : Enables Key assignment to each binary digit.
1 : Disables 4th 3rd 2nd 1st
0000
(binary)
Fnc02
0 to 7
key
0 : None
1 : Manual print command
2 : Hold
3 : Alternate switch
4 : Momentary switch
5 : Changing between gross weight and net weight
6 : Clear the tare weight
7 : Clear the zero value
NOTE: Clearing the zero value can be enabled
or disabled by CAL15.
0
Fnc03
1 to 3
Display
update rate
1 : 20 times/second
2 : 10 times/second
3 : 5 times/second
1
Fnc04
0 to 9
x display
0 : None
1 : Zero tracking in progress
2 : Alarm (Zero range setting error, over)
3 : key status
4: : Zero band
5 : HI output (Over the upper limit value)
6 : OK output (
Between upper and lower limit values)
7 : LO output (Below the lower limit value)
8 : User input 1
9 : User output 1
0
Fnc05
0 to 16
Digital filter
Selects a cutoff frequency.
0: None
1: 100.0 Hz 9: 7.0 Hz
2: 70.0 Hz 10: 5.6 Hz
3: 56.0 Hz 11: 4.0 Hz
4: 40.0 Hz 12: 2.8 Hz
5: 28.0 Hz 13: 2.0 Hz
6: 20.0 Hz 14: 1.4 Hz
7: 14.0 Hz 15: 1.0 Hz
8: 10.0 Hz 16: 0.7 Hz
15
Fnc07
1 to 3
Hold
1 : Normal hold
2 : Peak hold
3 : Averaging hold
1
Fnc08
-99999 to
99999
Set value of
near-zero
The reference value for near-zero.
Decimal point position is linked to CAL02.
10
Vista de página 44
1 2 ... 40 41 42 43 44 45 46 47 48 49 50 ... 55 56

Comentários a estes Manuais