AD-4430B - 33 -
Function No.
Parameter
and range
Function name Description
Default
value
Fnc09
1 to 2
Mass to be
compared with
near-zero
Item to be compared with near-zero.
1 : Gross weight
2 : Net weight
1
Fnc10
-99999 to 99999
Set value of
upper limit
Reference value for the upper limit.
Decimal point position is linked to CAL02.
10
Fnc11
-99999 to 99999
Set value of
lower limit
Reference value for the lower limit.
Decimal point position is linked to CAL02.
-10
Fnc12
1 to 2
Mass to be
compared with
upper and
lower limit
Item to be compared with the upper and lower limit.
1 : Gross weight
2 : Net weight
1
Fnc13
1 to 2
Output logic of
upper and
lower limit
Logic used when the result of the comparison with
the upper and lower limit is output
1 : Positive logic
2 : Negative logic
1
5.4.4. The List Of The Hold Function
Function No.
Parameter
and range
Function name Description
Default
value
HLd01
0.00 to 9.99
Averaging time
Time to perform the averaging. Unit : second.
0.00 is not averaged.
0.00
HLd02
0.00 to 9.99
Start wait time
Time to wait before commencing hold or averaging.
Unit : second.
0.00
HLd03
0 to 2
Condition of
automatic start
Condition to start the hold or averaging automatically.
0 : Do not use automatic start.
1 : Above the near-zero range, and stable.
2 : Above the near-zero range.
0
HLd04
0 to 1
Release due to
falling control
input
Release when control input is falling.
0 : Do not release
1 : Release
1
HLd05
0.00 to 9.99
Time release
Release after a set amount of time has passed.
Unit : second. 0.00 is not averaged.
0.00
HLd06
0 to 99999
Release due to
range of
fluctuation
Release when fluctuation from the holding value
exceeds a set value. The decimal point position is the
same as the setting of CAL02.
0
HLd07
0 to 1
Release in
near-zero range
Release when the weighing value is in the near-zero
range.
0 : Do not release.
1 : Release.
0
This hold function only works when Fnc07 (hold) is set to 2 (peak hold) or 3 (averaging hold).
This hold function has no function when Fnc07 (hold) is set to 1 (normal hold).
HLd01 (averaging time) works only when Fnc07 (hold) is set to 3 (averaging hold).
Vista de página 34