- 40 - AD-4430B
7.4. Initializing Parameters
The initialization mode restores the parameters of the default values to the FRAM.
Three types of initialization mode are available as shown below.
Initialization mode Display Description
RAM
initialization
RAM memory is initialized. The center of zero and tare
value will be restored to 0.
General functions
initialization
Data of the general functions stored in the FRAM are
reset to factor settings.
All data
initialization
All data stored in the FRAM are initialized. So calibration
data is included, calibrate the module again.
7.4.1. Initializing Mode For RAM And Function Parameters
Step 1 Press the key while pressing and holding the key ( + ) to
display . If returning to the weighing mode, press the key.
Step 2 Press the
key while pressing and holding the key (
+ ) to
display of the check mode.
Step 3 Select the initialization mode using the key. Press the key.
Step 4 Select an item to be initialized using the key. Press the key.
Step 5 Check that all status LED are blinking.
If performing the initialization, press the key for 3 seconds or more.
After initialization, all segments light and return to the weighing mode.
If canceling the initialization, press the key return to the weighing mode.
7.4.2. Initializing The Whole Data
Step 1 In the standby mode (While turning off the module),
Press + key to display .
If returning to the weighing mode, press the key.
Step 2 Press the key. to enter to calibration mode.
Step 3 Select all initialization mode using the key. Press the key.
Step 4 Check that all status LED are blinking.
If performing the initialization, press the key for 3 seconds or more.
After initialization, all segments light and return to the weighing mode.
If canceling the initialization, press the key return to the weighing mode.