NIBP
Standards
Creative™ NIBP meets ANSI/AAMI SP10
standard.
Technique
Oscillometric
Cuff Pressure Range
0mmHg - 300mmHg
Pressure Precision
±3mmHg
Cuff Inflation Time
<10 Seconds (typical adult cuff)
Overpressure Protection
Adult 300mmHg
Paediatric 240mmHg
Neonate 150mmHg
BP Measuring Range
Systolic pressure
Adult 40mmH-275mmHg
Paediatric 40mmHg-200mmHg
Neonatal 40mmHg- 135mmHg
Diastolic Pressure
Adult 10mmHg -210mmHg
Paediatric 10mmHg - 150mmHg
Neonatal 10mmHg - 95mmHg
Mean Arterial Pressure
Adult 20mmHg-230mmHg
Paediatric 20mmHg-165mmHg
Neonatal 20mmHg - 110mmHg
BP Measuring Accuracy
Maximal mean difference ±5mmHg
Maximal Standard Deviation
8mmHg
Measurement Modes
Adjustable Inflation Pressure, Manual, Auto,
STAT
TEMP (OPTIONAL)
Technology YSI 400 compatible
Measuring Range 25.0°C-45.0°C
Measuring Accuracy ±0.2°C
Responding Time <45s
SpO2
Transducer / Technology Dual-wave length LED
Creative™ LFC™ Patented Technology
SpO2 Measuring Range 35% - 100% (Adult and Paediatrics)
Low Perfusion Capability 0.4% - 5%
SpO2 Measuring Accuracy
ARMS < or = 2% for SpO2 range from 70-
100%
ISO 80601-2-61:2011 (supercedes ISO 9919)
PR Measuring Range 30bpm - 240bpm
PR Measuring Accuracy ±2bpm or ±2% whichever is greater
OTHER
Built In Power Supply 100-240VAC, 50/60Hz
Built In Rechargeable
Li-Ion Battery
11.1V 4400mAh
Display 3.5" color LCD and digital LEDs
Alarms
User adjustable Audible
& Visual Alarm
Communication Serial/Ethernet
OPTIONAL ACCESSORIES
Trolley Roll Stand, Wall Mount and EMS/Transport Case
Creative™ also supply compatible Central Monitoring options
PRINTER (OPTIONAL)
Type Built in thermal printer
Recording Speed 25 mm/s
Scroll Up
Select
Scroll Down
Change Screen
NIBP Start
Alarm Mute
Print
Power Saving Mode
NIBP Cycle Time Set