Vista de página 32
22
10
Power supply
display
Power supply
type
Displays the power currently used.
Displays the remaining battery charge.
11
Current time
display
Current time
Displays the current time.
5-4 Pause Display
When the home display is being shown, press the PAUSE key to pause the waveform.
Note
When the display is paused, press the HOME key or PAUSE key to return to the home display.
In the Pause display, press the ECG RECORD START/STOP key to print the ECG waveform of 16
seconds occurred right before the PAUSE key is pressed.
In the pause display, after about 5 minutes of inoperation, the home display will be indicated.