A&D SV-10A Manual de Instruções Página 42

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 41
40
7-3. How to Use
1 Attach the water jacket on the table along the guides.
Make sure that the water jacket is slid furthest along
the guides until stopped.
2 When the water jacket is used, be sure to raise or
remove the protector.
3 Use much care so that the sensor plates and the
temperature sensor do not come into contact with the
inner wall of the sample. If they are in contact, a
measurement error may occur.
4 Be sure to adjust the sample surface to the center of the narrow part of the sensor plates.
SV-1A SV-10A/SV-100A
Slide furthest
until sto
pp
ed
Guide
Guide
Protector
Raise or remove
the protector
Sam
p
le
surface
Sensor
protective
cover
Sample
surface
adjustment
position
Sample
surface
Surface
locator plate
(Shaded portion)
SV-10A/SV-100A
Sensor
protective
cove
r
S
amp
l
e
surface
adjustmen
t position
Vista de página 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 101 102

Comentários a estes Manuais

Sem comentários