WinCT-PlusInstructionManual(OP-08)Please see the separate document "Manual(for the AD8526)" if theAD8526(Ethernet Converter) is used.Ver.2.0
9Step 7 Enter the IP address, subnet mask, default gateway and click ”Next”. It may take a while to proceed to the next screen. Step 8
10 Step 9 Click “Finish”. Step 10 Quit after verifying that the IP address has been changed. Step 11 Write down the set IP address on the IP addres
11 5. Using Data Acquisition Software RsMulti5-1. Features of RsMultiThis software enables data acquisition from A&D’s electronic balances or pla
125-2. Setting RsMulti Start from the Start menu on the computer. Click Start, Program, A&D WinCT-Plus, RsMulti to start. To open
13 The Property window opens. Follow the steps below to set up: ①Provide a name for the connection, e.g. GH-252 (the name of the weighin
14 As shown below, GH-252 is registered in the Config window. For connection of multiple instruments, click the “Add” button to add the con
15 -Sending a send data command to the weighing instrument for data acquisition. Click the “Test” button to send the command s
16 Step 4 Quitting communication Click the “Stop” button. Step 5 Pasting data into an Excel file (Not applicable with Windows Vista)
175-4. Explanations about each function of the RsMulti (1) The RsMulti window Table ① Records received data. To change the title, doub
18 Manual / Repeat Manual or automatic command transmission can be selected. -When “Repeat” is selected, a command will be sent automatica
1Contents1. Introduction ---------------------------------------------------------------------------------------------------2. Specifications ------
19Command Data This function sends the same command to all the connected instruments. -When “All” is selected, the setup command wil
20(2) Config window To open, click “Open” from the Config menu in the RsMulti window. Displays the list of the connection settings. Table
21(3) Property window When Connection Name is selected (highlighted), click the “Property” button to display the Config window. It can be
22 -COM For COM connection, select this option button. Setup the RS232C communication settings. Port:Com: Com port (*1
236. License Agreement -It is prohibited to reproduce "WinCT-Plus" without our prior written consent. -Specifications of "WinCT-Plus&
247. Supplementary7-1. Using as a virtual COM port With attached software “ComPort Redirector”, the OP-08 connected to a LAN can be used as aCOM po
25 (3) Click the [Add IP] button. Enter the IP address and Port number of the OP-08 and click the OK button. In the default setti
26 (6) Click the [Save] button to save and then click the [Close] button to finish. (7) Now, communication with the OP-08 can be enabled by openi
277-2. Using a DHCP If a DHCP server is available for the network to which the OP-08 is connected, follow theprocedures described below to change the
28 Step 4. Check the “Obtain an IP address automatically” “Assign IP” buttons. Step 5. Check the DHCP box and then click the Next button.
21. IntroductionThis manual describes the operation to acquire data from a balance via an OP-08 Ethernetinterface by using data acquisition software “
29 Step 6 Click the Assign button. Step 7. Click the Finish button. Step 8. Click the Search button to display the IP address assigned by the DHCP.
308. Verify these items if communication problems occurIn the event that the communication fails even though you conducted your set up following thein
31OK button. Step 7 When the screen as shown below appears, click “Serial Settings”.
32 Step 8 The current settings for serial communication will be displayed. Confirm that the settingscorrespond with those shown in the table below .
33 Step 10 Click “Apply Settings” to apply the settings change. Please wait for completion. Step 11 Click “Connection” on the screen below.
34 Step 12 The screen below will be displayed. Confirm that the “Local Port” is “10001.” If thesetting is different, change the setting to “10001” an
35 Step 14 Click “Apply Settings” to apply the settings change. Please wait for completion. Step 15 Quit the DeviceInstaller.
32. Specifications OP-08 specificationsLAN interface10Base-T / 100Base-TX(Auto recognition)Shape of connector RJ45Protocol*1TCP/IP,UDP,ARP,ICMP,SNMP,
43.Installing Each Software If it is already installed, proceed to "4.Network Setting". For windows 2000, XP or Vista, login the compute
54. Network SettingTo implement a LAN connection, it is required to set the IP address, subnet mask, etc. for boththe computer and the OP-08. Please c
64-1. Setting a computer Set the IP address and the subnet mask. Open the Property window of TCP/IP, select “Use the following IP addres
74-2. Setting a weighing instrument and OP-08(1) Setting a weighing instrument Set the internal setting of the serial interface of the weighing ins
8 Step 5 Confirm that the IP address is selected. Click the Assign IP button. If the following settingis made without selecting the IP address, the i