13. APPENDIX
The table below lists the relation between the ISO 8655 requirements and the PT series pipette
accuracy testers.
ISO 8655 requirements
Maximum permissible error
Pipette
nominal
volume
*5
Accuracy Repeatability
Balance minimum
weighing value
(L) ±% ±L % L mg
1 5.0 0.05 5 0.05
2 4.0 0.08 2 0.04
5 2.5 0.125 1.5 0.075
10 1.2 0.12 0.8 0.08
0.001
20 1.0 0.2 0.5 0.1
50 1.0 0.5 0.4 0.2
100 0.8 0.8 0.3 0.3
0.01
200 0.8 1.6 0.3 0.6
500 0.8 4.0 0.3 1.5
1000 0.8 8.0 0.3 3.0
2000 0.8 16 0.3 6.0
5000 0.8 40 0.3 15
10000 0.6 60 0.3 30
0.1
Daily inspection, simplified verification*7
*5 The maximum volume selectable for adjustable volume pipettes
*6 The AD-4212B-PT can be used for the pipette volume range from 1 L to 10000 L.
*7 The minimum weighing value, 1 mg, approximately corresponds to 1 L. If a pipette volume is
1000 L, a test can be performed with a resolution of 0.1%. If 200 L, 0.5%.
NOTE Pipette accuracy testing is performed using a high-resolution valance. Make sure that
the measurement environment is free from vibration, drafts and air from air
conditioners. For details, refer to “3. PRECAUTIONS BEFORE USE”.
Pipette accuracy tester
corresponding to the requirements
The AD-4212B-PT*6
The AD-4212A-PT
FX-300i-PT