59
14-5 Memory for Upper/lower limit values
Upper/lower limit values stored in memory can be selected easily by pressing the MODE key.
Storage capacity for the 3-level comparison is 20 sets of upper/lower limit values (code
numbers C01-C20). "C01" is to store the upper and lower limit values that are set in the function
table.
The code number appears in the upper left of the display to label the value.
Using commands, the upper/lower limit values can be recalled and changed.
Notes
If the data memory function is used, the “Title block” and “End block” can not be output.
This function is not available when the BCD output (OP-01) is installed.
Setting the function table
1 Set the “Data memory ( data )” parameter to “4”.
Note
If “Clr” appears blinking in the upper left of the display, delete the data in memory.